Analisador de desfibrilador Phase 3 Semi novo
Destinatário: :
* Campos Obrigatórios
ou Cancelar
Tests Performed
| Defibrillator Energy, Charge Time and Cardioversion Automated External Defibrillator (AED) Performance ECG Monitor Performance Pacer Sensitivity, Refractory Periods and Noise Immunity Pacer Pulse Characteristics |
| Energy Measurement, General: | |
|---|---|
| Load resistance: | 50 ohms ±1%, non-inductive |
| ECG amplitude at defib pads: | 1 mV QRS |
| WAVEFORM (oscilloscope) Output | |
| High Range: | 1000:1 amplitude attenuation |
| Low Range: | 200:1 amplitude attenuation |
| Waveform Playback: | 200: 1 time base expansion |
| First Phase Pulse Width: | 0.1 - 58 msec |
| Second Phase Pulse Width: | 0.1 - 58 msec |
| Interphase Delay: | 0.1 - 58 msec |
| Tilt: | 0 - 99.9% |
| Modulation Frequency: | 1000 - 8000 Hz |
| Modulation Duty Cycle: | 0 - 99.9% |
| Test Pulse: | 46 Joules ±10% |
| Defibrillator High Range Energy Test: | |
|---|---|
| Energy Measurement: | 0.0 to 600.0 Joules (±1% ±2 LSD) |
| Voltage Measurement: | 0 to 5000 Volts (±1% ±2 LSD) |
| Current Measurement: | 0.0 to 100.0 Amps (±1% ±2 LSD) |
| Pulse Width Measurement: | 0.5 to 58.36 msec (±1% ±2 LSD) |
| Trigger Level: | 80 Volts |
| Playback Amplitude: | 1 mV per 1000 volts on Lead II; 1 mV per 2000 volts at defib pads |
| Test Pulse: | 126 Joules ±10% |
| Defibrillator Low Range Energy Test: | |
|---|---|
| Energy Measurement: | 0.0 to 50.0 Joules (±1% ±2 LSD) |
| Voltage Measurement: | 0 to 1000 Volts (±1% ±2 LSD) |
| Current Measurement: | 0.0 to 20.0 Amps (±1% ±2 LSD) |
| Pulse Width Measurement: | 0.5 to 58.36 msec. (±1% ±2 LSD) |
| Trigger Level: | 16 Volts |
| Playback Amplitude: | 1 mV per 200 volts on Lead II; 1 mV per 400 volts at defib paddles |
| Defibrillator Charge Time Test: | |
|---|---|
| Charge Time Measurement: 0.0 to 99.9 seconds (±1 LSD) |
| Defibrillator Cardioversion Test: | |
|---|---|
| Sync Delay Measurement: | -200 to +800 msec. (±1 LSD) |
| Delay Target: | 20 to 65 msec window when enabled |
| Sync Point: | Selectable, peak of ECG Q or R wave |
| AED Performance Test: | |
|---|---|
| Test Method: Verify AED shock advisory for specified arrhythmia |
| Pacemaker Pulse Test: | |
|---|---|
| Pulse Amplitude Measurement: | 4 to 250 milliamps, all loads (±1% ±1 LSD) |
| Pulse Rate Measurement: | 20 to 220 PPM (±1% ±1 LSD) |
| Pulse Width Measurement: | 0.5 to 58.36 msec. (±1% ±2 LSD) |
| Test Load Range: | 50 to 1600 ohms, in 50 ohm steps |
| WAVEFORM output: | 50 milliamps per volt, all loads |
| Measurement Methods: | Average, leading edge, trailing edge, peak |
| Test Pulse: | 145 mA ±10% |
| Pacemaker Noise Immunity Test: | |
|---|---|
| Test Waveform: | 50Hz or 60Hz sine wave |
| Noise Amplitude Range: | 0.18 to 10.00 mV peak-to-peak |
| Noise Amplitude Resolution: | 0.139 mV |
| Pacemaker Sensitivity Test: | |
|---|---|
| Test Waveform: | Square (SQR), Triangle (TRI) or Haversine (SSQ) pulse |
| Waveform Width: | 10, 25, 40, 100, or 200 msec. |
| Amplitude Range: | 0.00 to 3.00 mV peak |
| Pacemaker Refractory Period Test: | |
|---|---|
| Paced Refractory Period (PRP): | 50 to 750 msec. (±1 LSD) |
| Sensed Refractory Period (SRP): | 50 to 750 msec. (±1 LSD) |
ECG Simulator
| Performance Test Waveforms: | |
|---|---|
| DC Pulse, 4 seconds, Square Wave, 2 Hz, Triangle Wave, 2 Hz Sine Wave @ 0.1, 0.5, 10, 20, 40, 50, 60, 70, or 100 Hz |
| Normal Sinus Rhythm: | |
|---|---|
| 30, 60, 90, 120, 150, 180, 240 or 300 BPM |
| Cardioversion, Shock Advisory and AED Test Waveforms: | |
|---|---|
| Atrial Fibrillation, Coarse Atrial Fibrillation, Fine Asystole 1 (random, low-frequency baseline fluctuation) Asystole 2 (flat line/zero volts) Supraventricular Tachycardia (SVT-140) VTACH @ 140, 160, 190 BPM Torsades de Pointe @ 200 BPM Coarse Ventricular Fibrillation (CVF) / Fine Ventricular Fibrillation (FVF) Square Pulse 1ms to 60 BPM |
| Arrhythmia Simulations: | |
|---|---|
| Second Degree A-V Block, Premature Atrial Contraction (PAC), R-on-T PVC Right Bundle Branch Block (RBBB), Premature Ventricular Contraction (PVC) Multifocal PVC, Run of 5 PVC, Bigeminy, Trigeminy |
| Pacemaker Test Waveforms: | |
|---|---|
| SQR (square) Pacer Trigger, width = 2, 25, 40, 100 or 200 msec TRI (triangle) Pacer Trigger, width = 10, 25, 40, 100 or 200 msec SSQ (haversine) Pacer Trigger, width = 10, 25, 40, 100 or 200 msec |
| Performance Specifications: | |
|---|---|
| Output Level: | Selectable, 1 mV, 2 mV or 0.5 mV into ECG Lead II |
| Impedance: | 500 ohms (±0.2%), Amplitude: ±2% |
Non-Volatile Memory
| Data Capacity: | |
|---|---|
| 50 Test Records, 10 Defibrillator Waveform Records, 32 Autosequences |
| Test Record Content: | |
|---|---|
| Device ID, Time/date of test, Test type (Manual or Auto), Device type (defib. or AED) Up to 10 defibrillator energy tests (or 32 AED energy tests) 1 defibrillator charge time test Up to 4 defibrillator cardioversion tests Up to 12 ECG performance tests Up to 10 pacer pulse tests 1 pacer noise immunity test Up to 2 pacer sensitivity tests Up to 2 pacer refractory period tests |
| Interface: | |
|---|---|
| LCD (5.2" x 1.5"; 40 characters x 8 lines text; 240 x 64 pixel graphics) Defibrillator Input: Molex 42820-3212 Pacemaker Input: 2 x safety-style banana jack (red (+)/black (-)) ECG Simulator Outputs: 10 x safety banana jack (RA; RL; LA; LL; V1-V6) Defibrillator/Pacer Waveform Output: 1/8" mono phono jack High-Level ECG Output: 1/8" mono phono jack USB Port: Type "B", USB 1.1 or USB 2.0 compatible. Serial (RS-232) Port: DB9 Male, RS-232C, bi-dir, CTS handshaking, 9600 baud, 8-N-1 Keyboard Port: PS/2 (6-pin miniDIN female) |
| Power Supply: | |
|---|---|
| Internal 12.5V/1.4A-h NiCad, 24 hours of use between charges |
| Environment | |
|---|---|
| 15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II |
| Dimensions | |
|---|---|
| 9.5" W x 8" H x 5.5" D (24cm W x 20cm H x 14cm D) |
| Weight | |
|---|---|
| 3 lbs. (1.4 kg) |
All specifications subject to change without notice.